Osteointegration
The aim of this trial was to find the optimal surface treatment for MP1TM screws for best Osteointegration.
Three surface treatments were tested:
No treatment
Abrasion with SiC
Abrasion + Ta deposition by Plasma
Coating and abrasion were tried in order to promote the osteointegration into the bone.
Screws were implanted in rabbits for 4-6 weeks.
Results:
Tensile strength of plucking screws (N.cm) due to osteointegration atfter various implantation times and different surface treatments of MP1TM
Time implanted in rabits |
4 weeks |
6 Weeks |
No treatment |
13.08 |
17.93 |
Abrasion with SiC |
13.8 |
20.66 |
Abrasion + Ta deposition by Plasma |
16.5 |
24.5 |
Trial 2:
24 screws of 2 designs treated with Abrasion + Ta deposition by Plasma and 24 screws of Ti (control samples) were implanted in sheep bone for 4-12 weeks.
Results were excellent (see opinion letter)